Review of Scientific Instruments · 2012 · 73 citations · 27 references
Natural SciencesMaterials ScienceRadiation DetectionRadiation PhysicsSpectroscopyRadiation SpectroscopyX-ray DetectionX-ray DiffractionX-ray OpticsX-ray SpectroscopyHigh-resolution X-ray SpectroscopyModular DesignPosition-sensitive DetectionPosition-sensitive DetectorsSpectrometer DesignX-ray DiagnosticsX-ray Emission SpectroscopyX-ray Spectra
A complete in-vacuum curved-crystal x-ray emission spectrometer in Johansson geometry has been constructed for a 2-6 keV energy range with sub natural line-width energy resolution. The spectrometer is designed to measure x-ray emission induced by photon and charged particle impact on solid and gaseous targets. It works with a relatively large x-ray source placed inside the Rowland circle and employs position sensitive detection of diffracted x-rays. Its compact modular design enables fast and easy installation at a synchrotron or particle accelerator beamline. The paper presents main characteristics of the spectrometer and illustrates its capabilities by showing few selected experimental examples.
27
WIDTHS OF THE ATOMIC K–N7 LEVELS
T. Papp · Atomic Data and Nuclear Data Tables · 2001 · 754 citations