Review of Scientific Instruments · 2005 · 37 citations · 21 references
EngineeringMicroscopyMicrofabricationMechanicsMechanical EngineeringLateral Force CalibrationFrictional Force MicroscopeMicroscopy MethodScanning Force MicroscopyScanning Probe MicroscopyFriction ControlLateral ForceBiomedical EngineeringInstrumentationMedicineNanotribologyBiophysics
In this article, the construction and initial tests of a frictional force microscope are described. The instrument makes use of a microfabricated cantilever that allows one to independently measure the lateral forces in X and Y directions as well as the normal force. We use four fiber-optic interferometers to detect the motion of the sensor in three dimensions. The properties of our cantilevers allow easy and accurate normal and lateral force calibration, making it possible to measure the lateral force on a fully quantitative basis. First experiments on highly oriented pyrolytic graphite demonstrate that the microscope is capable of measuring lateral forces with a resolution down to 15pN.
21
Atomic-scale friction of a tungsten tip on a graphite surface
C. Mathew Mate, Gary M. McClelland, R. Erlandsson et al. · Physical Review Letters · 1987 · 1.7K citations