Review of Scientific Instruments · 2008 · 49 citations · 14 references
EngineeringBioengineeringMechanical EngineeringFluid DynamicsMicrofabricationRheologyBiofluid MechanicsCapillary RheologyMicrofluidic Device FabricationMicrofluidic DevicesMicrofluidic SystemsMicrofluidicsAnodic BondingMicrofluidic Device DesignMicrofluidic FabricationPressure DropShear RateShear StressSilicon MicrochannelsWall Shear Stress
We present the design and the process of fabrication of micromachined capillary on chip rheometers which have performed wall shear stress and shear rate measurements on silicon oil and ethanol-based nanofluids. The originality of these devices comes from the fact that local pressure drop measurements are performed inside the microchannels. Thus, the advantage over existing microviscometers is that they can be used with the fluid under test alone; no reference fluid nor posttreatment of the data are needed. Each on chip viscometer consists of anodically bonded silicon-Pyrex derivative microchannels equipped with local probes. The anodic bonding allows to reach relatively high pressure levels (up to approximately 10 bars) in the channels, and a broad range of shear stress and shear rate values is attainable. Dielectrophoretic and electrorheological effects can be highlighted by employing alternate microstripe electrodes patterned onto the inner side of the Pyrex wall.
14
The giant electrorheological effect in suspensions of nanoparticles
Weijia Wen, Xianxiang Huang, Shihe Yang et al. · Nature Materials · 2003 · 587 citations
Rheological behaviour of ethylene glycol based titania nanofluids
Haisheng Chen, Yulong Ding, Yurong He et al. · Chemical Physics Letters · 2007 · 494 citations