We have designed and built an imaging filter which can be attached to most stan- dard TEMs, and is capable of operating at primary energies of up to 400 keV. The filter uses a 90m agnetic sector prism, a piezoelectrically controlled energy-selecting slit, and 6 quadrupole and 5 sextupole lenses. It fully corrects second-order aberrations and distortions in images formed with electrons of selected energies, and it also produces second-order aberration-corrected spectra of vari- able dispersion. We show the first applications of the filter at 200 keV primary energy, which indicate that the filter will excel in chemical mapping and spectroscopy, in improving contrast of electron im- ages and diffraction patterns, and in making high resolution electron microscopy and diffraction more quantitative. We conclude the paper with a discussion about atomic resolution image formation using inelastically scattered electrons, and the relationship between energy-filtered diffraction patterns and images formed with electrons of the same energy.