New possibilities of the x-ray standing wave method in multiple diffraction of synchrotron radiation

A. Kazimirov, M. V. Kovalchuk, I. Yu. Kharitonov, L. V. Samoilova, Tetsuya Ishikawa, Shu Kikuta

Review of Scientific Instruments · 1992 · 35 citations · 12 references

Concepts

Abstract

Results of measurements of the reflected beam intensity and photoelectron yield under conditions of dynamical 111/220 diffraction of synchrotron x-ray radiation are presented. The experiments were carried out on beam lines 15C and 14B at the Photon Factory. Two possible multicrystal arrangements were used for the preprocessing of the incident beam. In the first one, a double-crystal Si (111) monochromator was used for monochromation and collimation of the x-ray beam in the vertical plane and a Si (220) channel-cut crystal was used for collimation in the horizontal plane. The second crystal arrangement is based on a (111)-oriented Si crystal adjusted for a six-beam Laue diffraction condition. The anomalously transmitted (μt/γ0 ≊ 40) beam was used as an incident beam in the x-ray standing wave measurements. The angular dependence of the total photoelectron yield was measured using a gas flow proportional counter.