Applied Physics Letters · 2008 · 84 citations · 12 references
EngineeringMaterials EngineeringMaterials ScienceHeat TransferMicrofabricationThin FilmsHeat Transfer In SolidsScanning Probe MicroscopySilicon MaterialsTemperature MeasurementThermal Properties Of SemiconductorsThin Film Thermal PropertiesHeat Transfer CharacterizationLateral ResolutionMicroscopySiliconSilicon DioxideSilicon NitrideThermal ConductivityVacuum Conditions
We present scanning thermal microscopy (SThM) measurements on a sample consisting of regions of 3nm thick HfO2 film and 2nm thick SiO2 on a silicon substrate. The experiments were preformed in high vacuum conditions using microfabricated silicon cantilevers with sharp heatable tips, facilitating the unprecedented achievement of a lateral SThM image resolution of 25nm. In addition, the heat transfer through the tip to the sample was investigated using approach curves and used to determine the thermal conductivity of the 3nm thick HfO2 layer.
12
Ed Swartz, R. O. Pohl · Reviews of Modern Physics · 1989 · 2.9K citations
Arun Majumdar · Annual Review of Materials Science · 1999 · 562 citations