EngineeringElectrical EngineeringPhotonicsSensorsSpectroscopyGas SensorsInfrared SpectroscopyLaser SpectroscopyNear-infrared SpectroscopyOptical InstrumentationOptical SpectroscopyDetection Limit AssessmentDetection Limit EstimationDetector CharacterizationDistributed Feedback LasersInfrared Gas SensorsInstrument Detection LimitWhite NoiseWhite Noise Excitation
In this Letter, a new method is proposed to further improve the detection sensitivity of the gas sensor based on off-axis integrated cavity spectroscopy (OA-ICOS) using a distributed feedback laser operating in the near-infrared at 1653 nm that relies on implementation of wavelength modulation and radio frequency (RF) white noise perturbation to diode laser current. More than sixfold improvement in detection limit is achieved for an RF noised wavelength-modulated OA-ICOS compared to the unperturbed conventional OA-ICOS approach.