Modeling and Calibration of a Novel One-Mirror Galvanometric Laser Scanner

Cheng-Yi Yu, Xiaobo Chen, Juntong Xi

Sensors · 2017 · 51 citations · 30 references

DOIFull text

Open access

Concepts

TL;DR

Laser stripe sensors are limited for collecting point clouds of object surfaces, motivating the development of a galvanometric laser scanner that sweeps a laser stripe across the surface using a single‑mirror galvanometer. The study aims to design and implement this one‑mirror galvanometric laser scanner to enable efficient surface point cloud acquisition. A novel, assumption‑free mathematical model is derived, and a model‑driven calibration procedure incorporating machining and assembly error effects is proposed, using a plane‑constraint method to extract numerous calibration points and evaluated for repeatability and accuracy on an automobile production line. Experimental results demonstrate that the calibration approach achieves measurement performance comparable to a look‑up table calibration method.

Abstract

A laser stripe sensor has limited application when a point cloud of geometric samples on the surface of the object needs to be collected, so a galvanometric laser scanner is designed by using a one-mirror galvanometer element as its mechanical device to drive the laser stripe to sweep along the object. A novel mathematical model is derived for the proposed galvanometer laser scanner without any position assumptions and then a model-driven calibration procedure is proposed. Compared with available model-driven approaches, the influence of machining and assembly errors is considered in the proposed model. Meanwhile, a plane-constraint-based approach is proposed to extract a large number of calibration points effectively and accurately to calibrate the galvanometric laser scanner. Repeatability and accuracy of the galvanometric laser scanner are evaluated on the automobile production line to verify the efficiency and accuracy of the proposed calibration method. Experimental results show that the proposed calibration approach yields similar measurement performance compared with a look-up table calibration method.

References

30