Analytical Chemistry · 2011 · 171 citations · 44 references
PharmacyPharmaceutical AnalysisBiocompatibilityFood AnalysisMicrofabricationCoating ProcessesCoating TechnologyCoatingsExtraction TechniquesMicrofabrication ProcessesSolid-phase ExtractionAutomated Solid-phase ExtractionBrush PaintingExtraction RecoveryHigh-throughput ExtractionMicroextractionPan CoatingPh StabilitySolid-phase Microextraction
Biocompatible C18-polyacrylonitrile (PAN) coating was used as the extraction phase for an automated 96-blade solid phase microextraction (SPME) system with thin-film geometry. Three different methods of coating preparation (dipping, brush painting, and spraying) were evaluated; the spraying method was optimum in terms of its stability and reusability. The high-throughput sample preparation was achieved by using a robotic autosampler that enabled simultaneous preparation of 96 samples in 96-well-plate format. The increased volume of the extraction phase of the C18-PAN thin film coating resulted in significant enhancement in the extraction recovery when compared with that of the C18-PAN rod fibers. Various factors, such as reusability, reproducibility, pH stability, and reliability of the coating were evaluated. The results showed that the C18-PAN 96-blade SPME coating presented good extraction recovery, long-term reusability, good reproducibility, and biocompatibility. The limits of detection and quantitation were in the ranges of 0.1-0.3 and 0.5-1 ng/mL for all four analytes.
44
Inge Bruheim, Xiaochuan Liu, Janusz Pawliszyn · Analytical Chemistry · 2003 · 462 citations
Sample Preparation: Quo Vadis?
Janusz Pawliszyn · Analytical Chemistry · 2003 · 283 citations · Full text