IEEE Journal of Selected Topics in Signal Processing · 2015 · 29 citations · 86 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsCell BiophysicsAtomic Force MicroscopyCellular ImagingImage ReconstructionScanning Probe MicroscopyAtomic Force Microscopy AnalysisComputational Reconstruction AlgorithmsImage InterpolationImage Reconstruction AlgorithmsMicroscopy ImagingReconstruction AlgorithmsReconstruction Quality Assessment
This paper provides a study of spatial undersampling in atomic force microscopy (AFM) imaging followed by different image reconstruction techniques based on sparse approximation as well as interpolation. The main reasons for using undersampling is that it reduces the path length and thereby the scanning time as well as the amount of interaction between the AFM probe and the specimen. It can easily be applied on conventional AFM hardware. Due to undersampling, it is necessary to subsequently process the acquired image in order to reconstruct an approximation of the image. Based on real AFM cell images, our simulations reveal that using a simple raster scanning pattern in combination with conventional image interpolation performs very well. Moreover, this combination enables a reduction by a factor 10 of the scanning time while retaining an average reconstruction quality around 36 dB PSNR on the tested cell images.