2013 · 24 citations · 5 references
EngineeringBiomedical EngineeringElectrical EngineeringAntennasMicroelectronicsSensorsImplantable MicrosystemsMems Energy HarvestingMems SensorsRadio-frequency Integrated CircuitsRf MemsTemperature SensorsChip AntennasEnergy HarvestingEnergy Harvesting SensorsEnergy Harvesting SystemsImplantable Pressure SensorsImplantable Temperature SensorsWireless Pressure SensorsWireless Sensing
For the first time a single chip implantable wireless sensor system for Intraocular Pressure Monitoring (IOPM) is presented. This system-on-chip (SoC) is battery-free and harvests energy from incoming RF signals. The chip is self-contained and does not require external components or bond wires to function. This 1.4mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">3</sup> SoC has separate 2.4GHz-transmit and 5.2GHz-receive antennas, an energy harvesting module, a temperature sensor, a 7-bit TIQ Flash ADC, a 4-bit RFID, a power management and control unit, and a VCO transmitter. The chip is fabricated in a standard 6-metal 0.18μm CMOS process and is designed to work with a post-processed MEMS pressure sensor. It consumes 513μW of peak power and when implanted inside the eye, it is designed to communicate with an external reader using on-off keying (OOK).
5
A cubic-millimeter energy-autonomous wireless intraocular pressure monitor
Gregory Chen, Hassan Ghaed, Razi Haque et al. · 2011 · 207 citations
Medicine, Ophthalmology, Surgery +14