American Journal of Public Health · 1992 · 15 citations · 22 references
Public HealthPreventive MedicineHealth AssessmentMethod ValidationStatisticsBioanalytical Method ValidationPopulation Health MeasurementPopulation ScreeningSampling MethodsScreening EvaluationDyslipidemia ScreeningFingerstick SamplingHigh-density Lipoprotein Cholesterol MeasurementLow-density Lipoprotein Cholesterol MeasurementPrecision AssessmentPrecision ScreeningReference Laboratory Standards
The precision, accuracy, and durability of the Reflotron were evaluated by the Massachusetts Model Systems for Blood Cholesterol Screening Program. Screenings were conducted in diverse community settings over 16 months. Fingerstick samples from 10,428 individuals were tested. None of the four analyzers met the 1992 standards for precision, although two met the 1992 standards for accuracy. More than 40% of Reflotron values differed from the reference laboratory values by upwards of 5%. More than 16% of individuals were misclassified in terms of their risk category. All four instruments malfunctioned during the project.
22
Archives of Internal Medicine · 1991 · 1.1K citations
Medicine, Cardiology, Dermatology +13
The National Cholesterol Education Program
James I. Cleeman · Clinics in Laboratory Medicine · 1989 · 713 citations
A Clinician's Guide to the Office Measurement of Cholesterol
James J. Burke · JAMA · 1988 · 63 citations
Bias and precision of cholesterol analysis by physician's office analyzers.
Thomas Koch, Uma Mehta, H Lee et al. · Clinical Chemistry · 1987 · 46 citations · Full text