Bias and precision of cholesterol analysis by physician's office analyzers.
46
Citations
6
References
1987
Year
Cholesterol AnalysisHealth PolicyCardiovascular DiseaseMedicineSerum PoolsHyperlipidemiaBiostatisticsWhole BloodDyslipidemiaClinical ChemistryPublic HealthLaboratory MedicineAtherosclerosisEpidemiologyNegative Bias
We studied the bias and precision of serum cholesterol analysis by physician's office analyzers. Total imprecision (CV range, %) for analysis of serum pools was: Abbott Vision 1.5%-1.9%; Ames Seralyzer 3.9%-4.5%; BMD Reflotron 2.3%-3.8%; Chrometrics Cholesterol Test System 2.3%-2.8%; Kodak DT-60 1.6%-2.2%. The Ames Seralyzer exhibited an excessive between-run component of variation. We collected, from 109 volunteers, samples of venous serum, heparin-treated whole blood, heparin-treated plasma, and fingerstick whole blood, and analyzed each type (where possible) with each system; serum was also analyzed in duplicate (by a proposed Reference Method) at the Centers for Disease Control. For assays with serum, the BMD Reflotron and Kodak DT-60 exhibited negative bias. All systems gave lower results for plasma and whole blood than for serum. All systems except the Kodak DT-60 were less precise for analysis of patients' sera than for analysis of serum pools; between-specimen variables may influence the results of these systems.