The crystallography beamline I711 at MAX II

Yngve Cerenius, Kenny Ståhl, L.A. Svensson, Thomas Ursby, Å. Oskarsson, J. Albertsson, Anders Liljas

Journal of Synchrotron Radiation · 2000 · 208 citations · 6 references

DOIFull text

Open access

Concepts

Abstract

A new X-ray crystallographic beamline is operational at the MAX II synchrotron in Lund. The beamline has been in regular use since August 1998 and is used both for macro- and small molecule diffraction as well as powder diffraction experiments. The radiation source is a 1.8 T multipole wiggler. The beam is focused vertically by a bendable mirror and horizontally by an asymmetrically cut Si(111) monochromator. The wavelength range is 0.8-1.55 A with a measured flux at 1 A of more than 10(11) photons s(-1) in 0.3 mm x 0.3 mm at the sample position. The station is currently equipped with a Mar345 imaging plate, a Bruker Smart 1000 area CCD detector and a Huber imaging-plate Guinier camera. An ADSC 210 area CCD detector is planned to be installed during 2000.

References

6