Japanese Journal of Applied Physics · 2001 · 36 citations · 14 references
EngineeringMaterials EngineeringPhysicsMicrofabricationOpticsRadiation MeasurementElectron Beam LithographyX-ray DetectionX-ray DiffractionX-ray OpticsBeam Spot SizeDiffraction-limited ResolutionFresnel Zone PlatesHard X-ray OpticsSpatial Resolution MeasurementX-ray Focusing OpticsX-ray Focusing SystemsX-ray Lithography
A hard X-ray microbeam with zone plate optics has been tested, and preliminary experiments on scanning microscopy have been performed. A Fresnel zone plate fabricated by the electron-beam lithography technique is used as an X-ray focusing device. The material of the zone structure is tantalum with thickness of about 1 µm, and the outermost zone width of the zone plate is 0.25 µm. A focused spot size measured by knife-edge scanning is 0.3 µm at an X-ray energy of 8 keV. Closer evaluation of the spatial resolution has been done by observing resolution test patterns in a scanning X-ray microscopy experiment, and fine patterns with a 0.2 µm-structure have been resolved.
14
A microscope for hard x rays based on parabolic compound refractive lenses
B. Lengeler, Christian G. Schroer, M. Richwin et al. · Applied Physics Letters · 1999 · 271 citations
Nanometer focusing of hard x rays by phase zone plates
W. Yun, Barry Lai, Zhonghou Cai et al. · Review of Scientific Instruments · 1999 · 222 citations