Radiation Protection Dosimetry · 2007 · 34 citations · 8 references
EngineeringElectrical EngineeringNuclear EngineeringRadiation DosimetryRadiation MeasurementSemiconductor DevicesDosimetry InstrumentationLight-emitting DiodesNeutron DetectionRadiation DetectorsSemiconductor DetectorsGallium ArsenideGallium Arsenide DetectorsSemiconductor Neutron Detectors
Displacement damage (DD) caused by fast neutrons in unbiased Gallium Arsenide (GaAs) light emitting diodes (LED) resulted in a reduction of the light output. On the other hand, a similar type of LED irradiated with gamma rays from a (60)Co source up to a dose level in excess of 1.0 kGy (1.0 x 10(5) rad) was found to show no significant drop of the light emission. This phenomenon was used to develop a low cost passive fluence monitor and kinetic energy released per unit mass dosemeter for accelerator-produced neutrons. These LED-dosemeters were used to assess the integrated fluence of photoneutrons, which were contaminated with a strong bremsstrahlung gamma-background generated by the 730 MeV superconducting electron linac driving the free electron laser in Hamburg (FLASH) at Deutsches Elektronen-Synchrotron. The applications of GaAs LED as a routine neutron fluence monitor and DD precursor for the electronic components located in high-energy accelerator environment are highlighted.