Contributions to precision and accuracy of monazite microprobe ages

Joseph M. Pyle

American Mineralogist · 2005 · 141 citations · 23 references

Concepts

Abstract

We examine the factors controlling accuracy and precision of monazite microprobe ages, using a JEOL 733 Superprobe equipped with 4 PET crystals, and both 1-atm gas flow Ar X-ray detectors and sealed Xe X-ray detectors. Multiple PET crystals allow for simultaneous determination of Pb concentration on up to 3 detectors, and the effects of different detector gases on spectral form can be addressed. Numerous factors in the X-ray production, detection, and counting sequence affect spectral form, including: choice of accelerating voltage, changes in d-spacing of the diffraction crystal, use of X-ray collimation slits, and type of detector gas. The energy difference between ArKα X-rays and XeLα X-rays results in, for 1-atm Ar detectors, escape peaks of second-order LREE L line X-rays that cannot be filtered using differential mode PHA. The second-order LREE energies are passed to the counter and produce, for a 140 mm Rowland circle, several problematic interferences in the Pb region of a monazite wavelength-dispersive (WD) spectrum. WD monazite spectra produced with Xe detectors are free from second-order LREE interferences in the Pb region; escape peaks of the secondorder LREE are filterable with differential mode PHA if Xe detectors are employed.