Life SciencesBiochemistryLipid BiochemistryMass Spectrometric AnalysisMetabolomicsBioanalytical Method ValidationLipid AnalysisLipid ChemistryLipidomicsAssay ReproducibilityLipid Class AnalysisLipid Composition AnalysisLipid ExtractionLipid ProfilingLipid QuantificationLong-term PerformanceMass Spectrometry-based Lipidomics
The stability of the lipid concentration levels in shotgun lipidomics analysis was tracked over a period of 3.5 years. Concentration levels in several lipid classes, such as phospholipids, were determined in human plasma lipid extracts. Impact of the following factors on the analysis was investigated: sample amount, internal standard amount, and sample dilution factor. Moreover, the reproducibility of lipid profiles obtained in both polarity modes was evaluated. Total number of samples analyzed was approximately 6800 and 7300 samples in negative and positive ion modes, respectively, out of which 610 and 639 instrument control samples were used in stability calculations. The assessed shotgun lipidomics approach showed to be remarkably robust and reproducible, requiring no batch corrections. Coefficients of variation (CVs) of lipid mean concentration measured with optimized analytical parameters were typically less than 15%. The high reproducibility indicated that no lipid degradation occurred during the monitored time period.
20
Lipidomics reveals a remarkable diversity of lipids in human plasma
Oswald Quehenberger, Aaron M. Armando, Alex Brown et al. · Journal of Lipid Research · 2010 · 1.3K citations · Full text
Britta Brügger, Gerhard Erben, Roger Sandhoff et al. · Proceedings of the National Academy of Sciences · 1997 · 882 citations