EngineeringChemical EngineeringMechanical EngineeringAnalytical ChemistryMicroelectromechanical SystemsSensorsGas ChromatographyMems DesignMems FabricationMems MaterialsMems SensorsMicroelectromechanical SensorsAlignment Quality EvaluationColumn SeparationFlame Ionization DetectionGas Separation MaterialsMicromachined SensorsPortable Analytical DevicesSurface Acoustic Wave Gas SensorsSurface Acoustic Wave Sensors
In the field of environmental measurement and security, a portable gas chromatograph (GC) is required for the on-site analysis of multiple hazardous gases. Although the gas separation column has been downsized using micro-electro-mechanical-systems (MEMS) technology, an MEMS column made of silicon and glass still does not have sufficient robustness and a sufficiently low fabrication cost for a portable GC. In this study, we fabricated a robust and inexpensive high-precision metal MEMS column by combining diffusion-bonded etched stainless-steel plates with alignment evaluation using acoustic microscopy. The separation performance was evaluated using a desktop GC with a flame ionization detector and we achieved the high separation performance comparable to the best silicon MEMS column fabricated using a dynamic coating method. As an application, we fabricated a palm-size surface acoustic wave (SAW) GC combining this column with a ball SAW sensor and succeeded in separating and detecting a mixture of volatile organic compounds.
17