High-accuracy differential interferometry for the investigation of phase objects

G. Pretzier, H. Jäger, Theo Neger

Measurement Science and Technology · 1993 · 31 citations · 14 references

Concepts

Abstract

The authors present an interferometric method for the investigation of phase objects which employs a Mach-Zehnder arrangement with objects being located outside. It functions as a differential interferometer with the practical advantage that the sensitivity and the carrier fringe system can be chosen arbitrarily and independently. The computerized evaluation is well adapted to the experimental set-up, which helps to achieve high accuracy of the results. For the case of radially symmetrical objects a numerical method is presented for calculating the refractive index distribution from data measured with the above mentioned type of differential interferometer. The results of a demonstration experiment are shown and further applications of this differential interferometer are discussed.

References

14