Journal of Vacuum Science and Technology · 1978 · 44 citations · 0 references
Natural SciencesMaterials EngineeringMaterials ScienceAnalytical ChemistryImpurity ProfilingThin FilmsAuger Electron SpectroscopyDepth EstimationChemical Composition DiagnosticsComposition-depth ProfilesDepth ProfilingFilm ThicknessFilm Thickness MeasurementImpuritiesMillingSample Resolution
It is demonstrated that the chemical composition of thick and multilayer insulating films can be easily depth profiled by Auger electron spectroscopy by laterally scanning an angle-lapped region that reveals the entire depth to be profiled. Conventional depth profiling of such structures is often difficult because of unstable sample charge-up conditions that occur when Auger measurements are made during ion milling. The angle-lap technique is nondestructive, in that it allows one to reexamine the entire profile whenever desired. It is thus particularly suitable for investigation of samples with initially unknown or trace impurities. Also, this technique is free of depth-resolution degradation and differential sputter yield effects usually associated with long-term ion milling of thick films. The absolute depth resolution of the angle-lap–Auger profiling method is independent of the film thickness and is equal to the electron beam diameter plus sample manipulator resolution divided by the angle-lap lateral magnification. An ultimate depth resolution of ∠35 Å would be possible using a 0.5-μm-diam beam, a 0.2° lap, and a 0.5-μm manipulator resolution. This is approaching the depth resolution associated with the finite escape depth of Auger electrons. Using a 10-μm electron beam and a 0.5° angle lap in the present study, a depth resolution of about 1000 Å was obtained, as predicted. The oxygen concentration in oxygen-rich polysilicon films measured by this technique agrees with that measured by an Electron Probe Microanalyzer. The Auger data also suggest that all oxygen atoms in the films are bonded to silicon in the form of SiO2.