Medical Physics · 1991 · 437 citations · 0 references
Natural SciencesMaterials EngineeringMaterials ScienceRadiation MeasurementRadiation PhysicsSpectroscopyX-ray Data AnalysisX-ray DetectionX-ray OpticsX-ray ProductionX-ray SpectroscopyX-ray DiagnosticsX-ray Emission SpectroscopyX-ray Source CharacterizationX-ray Spectral AnalysisX-ray Spectroscopy SimulationX-ray Spectrum FormationX-ray Target Materials
The model builds on prior work by extending it in two significant ways. The study presents a semiempirical model for generating tungsten target x‑ray spectra. The model assumes bremsstrahlung and characteristic x‑ray production occur at varying depths within the target and optimizes parameters using nonlinear least‑squares fitting to experimental spectra. The model achieves good agreement between calculated and measured x‑ray tube spectra across various target angles and tube potentials, outperforming previously published models.
A semiempirical model for generating tungsten target x‐ray spectra is presented. This model extends earlier work in two significant areas. First, both bremsstrahlung and characteristic x‐ray production are assumed to occur at varying depths within the target. Second, optimal parameters for the model were determined from experimental spectra utilizing nonlinear least‐squares techniques. As a result, good agreement is obtained between calculated and measured x‐ray tube spectra and output for different target angles and a wide range of x‐ray tube potentials. Such is not the case with previously published models.