Review of Scientific Instruments · 1995 · 100 citations · 6 references
EngineeringBiophysicsInstrumentationOpticsSensorsAtomic Force MicroscopyCalibrationForce SensingForce SensorsOptical InstrumentationOptical InstrumentsAtomic Force Microscopy CalibrationCantilever Deflection DetectionDetector Sensitivity CalibrationDisplacementOptical DetectionOptical Force SensorsSensitivity CalibrationVoltage
Accurate force determinations in atomic force microscopy require the precise measurement of cantilever deflections. A limiting factor in making these measurements is the calibration of the optical lever detection system, particularly when the tip cannot be pressed against a hard surface. Here we show that, for a given instrument, a fixed displacement of the photodiode detector produces a change in the detector voltage that correlates strongly with optical lever sensitivity. This provides a simple method for optical lever sensitivity calibration not requiring contact of the tip with any surface. The data also suggest that differences in the shape of the laser spot on the photodiode are a major source of variability in optical lever sensitivity.
6
G. Binnig, C. F. Quate, Ch. Gerber · Physical Review Letters · 1986 · 14.4K citations · Full text
Calibration of atomic-force microscope tips
Jeffrey L. Hutter, John Bechhoefer · Review of Scientific Instruments · 1993 · 4.2K citations
Novel optical approach to atomic force microscopy
Gerhard Meyer, Nabil M. Amer · Applied Physics Letters · 1988 · 1.3K citations