EngineeringMaterials EngineeringBiomedical ImagingDiagnostic ImagingRadiation TransportRadiographic ImagingSuperconducting Thin FilmsX-ray DetectionX-ray DetectorsX-ray OpticsJosephson TunnelingJunction ModelingMicroscope ImagingPenetration DepthSynchrotron Radiation DetectionTunnel JunctionsX-ray Imaging PhysicsX-ray Imaging Systems
Imaging diagnosis of superconducting tunnel junction x-ray detectors has been performed by an apparatus called the low-temperature scanning synchrotron microscope (LTSSM) using an x-ray microbeam with a diameter of 5–10 μm originated from synchrotron radiation. Quasiparallel intense synchrotron radiation enables one to obtain the full two-dimensional images of junctions with dimensions of 200×200 μm2 in about 1 h. The LTSSM results indicate that the standard quasiparticle diffusion and edge loss model for the spatial distribution of the junction response to x rays is evidently inadequate for intermediate or large junctions (with respect to a Josephson penetration depth). On this basis, it is argued that the models proposed for the signal creation and loss mechanism should be reconsidered.