Palm NMR and 1-Chip NMR

Nan Sun, Tae‐Jong Yoon, Hakho Lee, William Andress, Ralph Weissleder, Donhee Ham

IEEE Journal of Solid-State Circuits · 2010 · 152 citations · 15 references

Concepts

TL;DR

The paper introduces two new NMR systems that dramatically reduce size and enable lab‑on‑a‑chip operation. The systems employ a CMOS RF transceiver integrated with NMR physics, with one design adding a sample coil for direct chip‑sample interfacing. The 0.1‑kg palm‑sized system is the smallest NMR ever built, offering 1200× size and weight reductions and 150× higher sensitivity, while the chip‑integrated version detects biomolecules such as avidin, hCG, and bladder cancer cells.

Abstract

In our earlier work, we developed a 2-kg NMR system, which was 60 X lighter, 40× smaller, yet 60× more spin-mass sensitive than a 120-kg state-of-the-art commercial benchtop system. Here we report on two new nuclear magnetic resonance (NMR) systems that represent further orders-of-magnitude size reduction and lab-on-a-chip capability. The first system, which weighs 0.1 kg and can be held in the palm of a hand, is the smallest NMR system ever built, and is 1200× lighter, 1200× smaller, yet 150× more spin-mass sensitive than the commercial system. It is enabled by combining the physics of NMR with a CMOS RF transceiver. The second system, which even integrates a sample coil, directly interfaces the CMOS chip with a sample for lab-on-a-chip operation. The two systems detect biological objects such as avidin, human chorionic gonadotropin, and human bladder cancer cells.

References

15