Femtotesla atomic magnetometry in a microfabricated vapor cell

W. C. Griffith, Svenja Knappe, John Kitching

Optics Express · 2010 · 185 citations · 20 references

DOIFull text

Open access

Concepts

Abstract

We describe an optically pumped 87Rb magnetometer with 5 fT/Hz(1/2) sensitivity when operated in the spin-exchange relaxation free (SERF) regime. The magnetometer uses a microfabricated vapor cell consisting of a cavity etched in a 1 mm thick silicon wafer with anodically bonded Pyrex windows. The measurement volume of the magnetometer is 1 mm3, defined by the overlap region of a circularly polarized pump laser and a linearly polarized probe laser, both operated near 795 nm. Sensitivity limitations unique to the use of microfabricated cells are discussed.

References