Open Journal of Applied Biosensor · 2013 · 18 citations · 24 references
EngineeringBioengineeringBiotechnologyBioinstrumentationBiosensorsSensorsBiological SensorsBiosensors For Immune DetectionEnzyme-linked Immunosorbent AssayImmunoassay AnalysisMycotoxinsPiezoelectric BiosensorsGold ElectrodesImmunosensorsPiezoelectric ImmunosensorsPiezoelectric SensorsQuartz Crystal MicrobalanceQuartz Crystal Microbalance BiosensorsQuartz Crystal Microbalance ImmunosensorsQuartz Crystal Microbalance Sensors
In this study, we have used a direct immunoassay where the simple binding between antigen and an antibody is detected. Immunoassays were performed in a drop system, monitoring the frequency decrease of the quartz-crystal microbalance device because of mass increasing during immunoreaction. The QCM sensor was coated on both sides by gold electrodes, only one side of the crystal (liquid side) was in contact with the solution; the other side (contact side) was always dry. We tested a piezoelectric immunosensor for aflatoxin B1 (AFLA-B1) mycotoxin detection through the immobilization of DSP-anti-AFLAB1 antibody (AFLA-B1-Ab anti AFLAB1) on gold-coated quartz crystals (AT-cut/5 MHz). The DSP (3,3'-Dithiodipropionic-acid-di-N-hydroxysuccinimide ester) was used for the covalent attachment of the proteins. The piezoelectric crystal electrodes were pretreated by DSP for 15 min, rinsed with water and dried in a gentle flow of nitrogen gas. Then the DSP-coated crystals were installed in a sample holder and exposed to the anti-AFLAB1 antibody and to the AFLA-BI. Frequency and resistance shifts (f and R) were measured simultaneously. f versus AFLA-BI concentrations in the range of 0.5 -10 ppb exhibited a perfect linear correlation with a coefficient of above 0.998.
24
Anna Yu. Kolosova, Won‐Bo Shim, Zhengyou Yang et al. · Analytical and Bioanalytical Chemistry · 2005 · 226 citations