Performance of integrated thermopile vacuum sensors

A.W. van Herwaarden, P.M. Sarro

Journal of Physics E Scientific Instruments · 1988 · 27 citations · 7 references

Concepts

Abstract

A quantitative insight is given into the performance of integrated thermopile vacuum sensors as a vacuum gauge. After a short resume of their operation principle three pressure ranges are discussed. The zero repeatability is shown to be approximately 10 mPa. In the optimum pressure range of 1-100 Pa the error of prediction of the sensors is 2-3% while the upper pressure limit of the sensors is of the order of 5-50 kPa. A comparison with conventional Pirani gauges is made.