Proceedings of the Physical Society · 1959 · 62 citations · 13 references
Natural SciencesChemistryAnalytical ChemistrySpectroscopic AnalysisSpectroscopyRadiation SpectroscopyX-ray Data AnalysisX-ray OpticsX-ray SpectroscopyCalibration ProceduresMicroanalysisSpectral CharacterizationX-ray DiagnosticsX-ray Emission LinesX-ray Emission SpectroscopyX-ray SpectraX-ray Spectral Analysis
Three different pulse analysis methods are described by which the intensities of several closely spaced spectral lines, as detected by a proportional counter, may be determined quickly and unambiguously. The techniques are applicable where each significant line occurring in the spectrum can be obtained separately for calibration procedures; X-ray emission microanalysis, for example, fulfils this condition. The two methods which have been tested gave absolute errors in the order of a few per cent in analysing various mixtures of copper and nickel. An important application of such techniques is in the analysis of characteristic X-rays having long wavelengths and low intensity, since maintenance of an adequate counting rate under these conditions is difficult if a crystal or ruled grating spectrometer is used.
13