Publication | Closed Access
Evaluation of the primary frequency standard NPL-CsF1
79
Citations
17
References
2005
Year
Microwave SpectroscopyEngineeringPhysicsMeasurementNatural SciencesSpectroscopyApplied PhysicsAtomic PhysicsCold AtomsSystematic Frequency ShiftsInstrumentationTimefrequency AnalysisFrequency ControlPrimary FrequencySignal ProcessingFrequency ShiftStatisticsElectromagnetic Compatibility
A new caesium fountain frequency standard (NPL-CsF1) at the National Physical Laboratory is described. Procedures for evaluation of the systematic frequency shifts are presented. The NPL-CsF1 has a short-term stability σy(τ) = 1.4 × 10−13τ−1/2, predominantly due to the local oscillator phase noise. The accuracy of 1 part in 1015 is limited by the uncertainty of the frequency shift due to collisions between cold atoms.
| Year | Citations | |
|---|---|---|
Page 1
Page 1