Applied Physics Letters · 2004 · 55 citations · 9 references
EngineeringDifferential Scanning CalorimetryMicroscopyAnalytical MicrosystemsBiomedical EngineeringChemistryCalorimetry MeasurementsBiosensing SystemsSensitive Ac NanocalorimeterCalorimetryAnalytical ChemistryBioimagingThermodynamicsNanometrologyNanosensorMicrofluidicsBiophysicsThermoanalytical MethodNanofluidicsBiopolymersCalorimetric MethodNano ScaleMicrofabricationBiomedical DiagnosticsAc CalorimeterUltrasensitive Ac NanocalorimeterMedicineThermophysical Property
We developed an ultrasensitive ac nanocalorimeter for use with biological liquids by means of microfabrication technologies. The volume of the cell measurement is only 5 μL. This nanocalorimeter, tested on deionized water, allows the measurement of heat capacity variation above ±150 nJ/K (resolution of ±5×10−6) with an operating temperature ranging from −20 to 120 °C and a stability of 100 μK. Its use is demonstrated on diluted lysosyme solution measured at the frequency of 3 Hz. At 3 Hz, this ac calorimeter gives only the variation of the heat capacity during the denaturation, which allows complementary thermodynamic investigations as regards to classical differential scanning calorimetry measurements.
9
Steady-State, ac-Temperature Calorimetry
Paul F. Sullivan, G. M. Seidel · Physical Review · 1968 · 745 citations
Engineering, Differential Scanning Calorimetry, Measurement +21