Journal of Applied Crystallography · 1997 · 19 citations · 1 references
Natural SciencesChemistryAnalytical ChemistryRadiation MeasurementSpectroscopyGas-filled DetectorsScattering TheorySmall-angle ScatteringX-ray CrystallographyX-ray Data AnalysisX-ray DetectorsGas DetectionGas Detector PhysicsGas Pressure EffectsGuinier AnalysisPosition-sensitive DetectionPosition-sensitive DetectorsResponse Function ModelingSmall-angle X-ray Scattering
The response of gas-filled position-sensitive detectors consists not only of the main Gaussian response (FWHM ~0.1 mm) but also of a linear exponential term measurable up to 7 mm from the excitation, superimposed on an incoherent background. This effect depends on gas type and pressure and is due to fluorescence of the detection gas. The consequences of this effect on some standard experimental methods used in small-angle X-ray scattering (SAXS), i.e. Porod limit and Guinier radii determination, are investigated.