Benefits of two-dimensional detectors for synchrotron X-ray diffraction studies of thin film mechanical behavior

Guillaume Géandier, P.-O. Renault, Simon J. Teat, Éric Le Bourhis, B. Lamongie, P. Goudeau

Journal of Applied Crystallography · 2008 · 19 citations · 35 references

Concepts

Abstract

Performing a complete in situ mechanical property analysis of polycrystalline thin films using X-ray diffraction is time consuming with most standard diffraction beamlines at synchrotron facilities and not realistic with laboratory diffractometers. Two-dimensional detection is shown to enable relatively fast and reliable X-ray strain measurements during in situ tensile testing of gold films deposited on polyimide substrates. Advantages and drawbacks in the use of two-dimensional detectors for this type of analysis are discussed for two commonly used geometries: reflection and transmission.

References

35