Journal of Separation Science · 2011 · 24 citations · 19 references
The quantitative performance of a simple home-built preparative gas chromatography (prep-GC) arrangement was tested, incorporating a micro-fluidic Deans switch, with collection of the target compound in a deactivated uncoated capillary tube. Repeat injections of a standard solution and peppermint sample were made into the prep-GC instrument. Individual compounds were eluted from the trapping capillary, and made up to constant volume. Chloronaphthalene internal standard was added in some cases. Recovered samples were quantitatively assayed by using GC-MS. Calibration linearity of GC-MS for menthol standard area response against number of injections (2-20 repeat injections) was excellent, giving R(2) of 0.996. For peppermint, menthol correlation over 2-20 repeated injections was 0.998 for menthol area ratio (versus IS) data. Menthone calibration for peppermint gave an R(2) of 0.972. (1) H NMR spectroscopy was conducted on both menthol and menthone. Good correspondence with reference spectra was obtained. About 80 μg of isolated menthol and menthone solute was collected over a sequence of 80 repeat injections from the peppermint sample, as assayed by 600 MHz (1) H NMR analysis (∼100% recovery for menthol from peppermint). A procedure is proposed for prediction of number of injections required to acquire sufficient material for NMR detection.
19
A new technique for heart cutting in gas chromatography [1]
D. R. Deans · Chromatographia · 1968 · 340 citations
W. Bertsch · Journal of High Resolution Chromatography · 1999 · 193 citations
Engineering, Supercritical Fluid Chromatography, Biomedical Engineering +17
Identification of the Sex Pheromone of the German Cockroach, <i>Blattella germanica</i>
Satoshi Nojima, Coby Schal, Francis X. Webster et al. · Science · 2005 · 118 citations