Review of Scientific Instruments · 2005 · 72 citations · 13 references
Natural SciencesBiomedical EngineeringElectrical EngineeringBiomedical ImagingDiagnostic ImagingSpectroscopyElectron Paramagnetic ResonanceElectron Spin Resonance SpectroscopyFourier Transform SpectroscopyMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance SpectroscopyDirect DetectionElectron Paramagnetic Resonance ImagingFourier Transform Nuclear Magnetic Resonance SpectroscopyFree Induction DecayIn Vivo StudiesMagnetic Resonance Imaging AcquisitionPulsed Electron Paramagnetic ResonancePulsed Electron Spin Resonance
The application of direct time-locked subsampling (TLSS) to Fourier transform electron paramagnetic resonance (FT-EPR) spectroscopy at radio frequencies (rf) is described. With conventional FT-EPR spectroscopy, the high Larmor frequencies (L(f)) often necessitate the use of intermediate frequency (IF) stages to down convert the received free induction decay (FID) signal to a frequency that can be acquired with common data acquisition technology. However, our research focuses on in vivo studies, and consequently utilizes a FT-EPR system with a L(f) of 300 MHz. This relatively low frequency L(f), in conjunction with the advent of bandpass sampling analog-to-digital conversion and signal processing technologies, has enabled us to omit the IF stage in our FT-EPR system. With this in mind, TLSS techniques have been developed to directly sample the 300 MHz FID signal at a sampling rate of 80 MHz providing a signal bandwidth of 20 MHz. The required modifications to the data acquisition and processing system specific to this application are described. Custom software developed to control the EPR system setup, acquire the signals, and post process the data, is outlined. Data was acquired applying both coherent averaging and stochastic excitation sequences. The results of these experiments demonstrate digital down conversion of the 300 MHz FID signal to quadrature baseband. Direct FID TLSS eliminates many noise sources common in EPR systems employing traditional analog receiver techniques, such as the IF mixer stage in single channel systems, and the quadrature baseband mixer stage in dual channel systems.
13
Nallathamby Devasahayam, Sankaran Subramanian, Ramachandran Murugesan et al. · Journal of Magnetic Resonance · 2000 · 68 citations
Engineering, Biomedical Engineering, Electrical Engineering +14
Noninvasive in vivo oximetric imaging by radiofrequency FT EPR
Sankaran Subramanian, Ken‐ichi Yamada, Akira Irie et al. · Magnetic Resonance in Medicine · 2002 · 60 citations · Full text