X-Ray Spectrometry · 1992 · 16 citations · 8 references
Natural SciencesChemistryPhysicsRadiation DetectionRadiation PhysicsSpectroscopyRadiation SpectroscopyX-ray Data AnalysisX-ray DetectionX-ray DetectorsX-ray SpectroscopyX-ray Emission LinesX-ray Line EmissionX-ray SpectraX-ray Spectroscopy Detectors
Abstract If the lineshape response of a Si(Li) detector is measured using monoenergetic x‐rays provided by a monochromator, it must be convoluted by a Lorentzian of appropriate width in order to represent the lineshape encountered in directly measured x‐ray spectra. It is shown that the Lorentzian component is responsible for a significant fraction of low‐energy peak tailing, a phenomenon usually attributed to detector imperfections.