Minimizing Carry-Over in an Online Pepsin Digestion System used for the H/D Exchange Mass Spectrometric Analysis of an IgG1 Monoclonal Antibody

Ranajoy Majumdar, Prakash Manikwar, John M. Hickey, Jayant Arora, C. Russell Middaugh, David B. Volkin, David D. Weis

Journal of the American Society for Mass Spectrometry · 2012 · 80 citations · 25 references

Concepts

Abstract

Chromatographic carry-over can severely distort measurements of amide H/D exchange in proteins analyzed by LC/MS. In this work, we explored the origin of carry-over in the online digestion of an IgG1 monoclonal antibody using an immobilized pepsin column under quenched H/D exchange conditions (pH 2.5, 0 °C). From a consensus list of 169 different peptides consistently detected during digestion of this large, ~150 kDa protein, approximately 30% of the peptic peptides exhibited carry-over. The majority of carry-over originates from the online digestion. Carry-over can be substantially decreased by washing the online digestion flow-path and pepsin column with two wash cocktails: [acetonitrile (5%)/isopropanol (5%)/acetic acid (20%) in water] and [2 M guanidine hydrochloride in 100 mM phosphate buffer pH 2.5]. Extended use of this two-step washing procedure does not adversely affect the specificity or activity of the immobilized pepsin column. The results suggest that although the mechanism of carry-over appears to be chemical in nature, and not hydrodynamic, carry-over cannot be attributed to a single factor such as mass, abundance, pI, or hydrophobicity of the peptides.

References

25