Journal of Biomedical Materials Research · 1976 · 292 citations · 7 references
Natural SciencesBioengineeringBiomedical EngineeringBiocompatibilityBiomaterialsPolymer MaterialsBiomaterials For Tissue EngineeringImplant MaterialsImplantable BiomaterialsPolymeric BiomaterialsEnzyme ActivityPolyethylene ImplantsPolymer BiocompatibilityPolymer ImplantsPolymeric Implant MaterialsPolymeric ImplantsSilicone ElastomersSilicone Implants
The study aimed to show that implant sample shape must be standardized to reliably assess polymer histotoxicity. Six medical‑grade polymers were fabricated into circular, triangular, and pentagonal rods and implanted into rat gluteal muscle for 14 days, with tissue response quantified by lysosomal acid phosphatase activity measured microspectrophotometrically. Triangular implants induced the strongest lysosomal activity, pentagons intermediate, and circular rods the weakest, underscoring the necessity of a standardized implant shape for comparative polymer toxicity studies.
Abstract Studies were designed to demonstrate the need for standardization of shape of samples used as implants to evaluate histotoxicity of polymer materials. Six medical‐grade polymers (polypropylene, polyethylene, polyurethane, silicone rubber, poly(vinyl chloride), and Teflon) were extruded as rods with circular‐, triangular‐, and pentagonal‐shaped crosss ections, and were implanted in rat gluteal muscles for 14 days. Evaluation of the tissue response was assessed by quantitating cellular lysosomal acid phosphatase enzyme activity by using microspectrophotometry. All triangular‐shaped implants showed the highest enzyme activity and cellular response; pentagon shapes showed less, and circular rods showed the lowest activity. The results demonstrate the need for standard sample shape for valid comparative studies of tissue response to implanted polymers.
7