Development of µGC (micro gas chromatography) with high performance micromachined chip column

Masanori Nishino, Yusuke Takemori, Satoshi Matsuoka, Masaki Kanai, T. Nishimoto, Masahito Ueda, Kyoichi Komori

IEEJ Transactions on Electrical and Electronic Engineering · 2009 · 24 citations · 9 references

Concepts

Abstract

Abstract A micro gas chromatography (µGC) instrument applying a high performance chip column fabricated on a silicon wafer was developed. Experimental results of the chip column and protyping of a µGC instrument are described. Approximately 35 000 theoretical plates were generated with the chip column coated liquid phase (5% phenyl‐/95% dimethyl‐polysiloxane). The theoretical plates of the chip column were close to those of the capillary column. Experimental minimum height equivalent to a theoretical plate (HETP, H min ) of the chip column was 1.2 times higher than the calculated H min . A prototype µGC applying the chip column was developed. The µGC generated approximately 35 000 theoretical plates, similar to the theoretical plates obtained by a commercial GC instrument. Copyright © 2009 Institute of Electrical Engineers of Japan. Published by John Wiley & Sons, Inc.