Optics Letters · 2008 · 136 citations · 14 references
Natural SciencesBiophysicsPhysicsBiophotonicsOpticsPhotonicsCellular ImagingDiffractionFluorescence MicroscopyLight MicroscopyOptical ImagingWave DiffractionLateral ResolutionLateral Resolution ImprovementPenetration DepthPenetration Depth DependenceStructured Illumination MicroscopyTotal Internal ReflectionTotal Internal Reflection Fluorescence MicroscopyWide-field Microscopy
In wide-field fluorescence microscopy, illuminating the specimen with evanescent standing waves increases lateral resolution more than twofold. We report a versatile setup for standing-wave illumination in total internal reflection fluorescence microscopy. An adjustable diffraction grating written on a phase-only spatial light modulator controls the illumination field. Selecting appropriate diffraction orders and displaying a sheared (tilted) diffraction grating allows one to tune the penetration depth in very fine steps. The setup achieves 91 nm lateral resolution for green emission.
14
Imaging Intracellular Fluorescent Proteins at Nanometer Resolution
Eric Betzig, George H. Patterson, Rachid Sougrat et al. · Science · 2006 · 8.9K citations