American Journal of Clinical Pathology · 1989 · 24 citations · 12 references
MedicineHematologyBiosensorsHematology PhysiologyHemostasisPlatelet ActivationPlatelet AdhesionPlatelet AggregationPlatelet FunctionPlatelet PhysiologyAdenosine Diphosphate-induced Platelet AggregationCollagen-induced Platelet AggregationElectrical ImpedanceImpedance AggregometryPlatelet Activation AssaysPlatelet Aggregometry
Platelet function in platelet-rich plasma may be evaluated using an optical or electrical impedance method. Platelet function in whole blood may be studied using an impedance or particle counter method. A comparison of these techniques in whole blood has not been performed. Twenty-four healthy subjects were studied using both techniques over a range of concentrations of collagen and adenosine diphosphate (ADP). Although broad similarities exist, discordant patterns occur in individual subjects. The particle counter method is capable of detecting small aggregates and thus is more sensitive. The impedance aggregometer detects only larger aggregate formation. These techniques are complementary rather than competitive and may have considerable potential in profiling platelet function.
12
The aggregation of blood platelets
G. V. R. Born, Michael Cross · The Journal of Physiology · 1963 · 2.1K citations
Medicine, Cell Biology, Hematology +12
IMPROVED AORTOCORONARY BYPASS PATENCY BY LOW-DOSE ASPIRIN (100 mg DAILY)
Reinhard Lorenz, Michael A. Weber, J. Kotzur et al. · The Lancet · 1984 · 321 citations