The results of this study support the manufacturer's recommendations to store POCT glucose test strips in their original vial, capped, and at room temperature, though refrigeration may offer an alternative storage environment with acceptable stability. As compliance with testing, quality control, and storage instructions is often an issue with POCT, the manufacturers of these systems for blood glucose measurement should design storage systems that allow the patient to store the glucose meter and the reagent strips in the same location. Manufacturers may also need to consider designing storage systems that are more portable, knowing that patients must take the glucose meters and test strips with them when they travel. Roche's Accu-Chek Compact system is an example of such a design. The glucose test strips are incorporated into a drum that is stored in the Accu-Chek meter itself. When a patient performs a fingerstick blood glucose measurement, the drum advances to move a test strip outside the meter. When the test is complete, the test strip is ejected for disposal. Future studies to clarify the effect of adverse storage conditions, particularly refrigeration, on the integrity of POCT test systems and reagent strips is warranted with currently marketed brands.