High Pressure Liquid Chromatographic Determination of Carbohydrates in Food Products: Evaluation of Method

Jonathan W De Vries, John C Heroff, David C. Egberg

Journal of AOAC INTERNATIONAL · 1979 · 20 citations · 0 references

DOIFull text

Open access

Concepts

Abstract

A high pressure liquid chromatographic (HPLC) method for determining 5 common carbohydrates in food products was evaluated. Reproducibility data were generated showing a relative standard deviation of 2.8%. Recovery studies on a variety of foods gave an average recovery of 98.8%. The HPLC data for 3 varieties of ready-to-eat cereals were compared with data from 4 independent laboratories using current AOAC chemical methods. The HPLC mean values differed from the chemical mean values by 3.2%.