Short-term stability of a microcell optical reference based on the Rb atom two-photon transition at 778  nm

Martin Callejo, Andrei Mursa, Rémy Vicarini, Emmanuel Klinger, Quentin A. A. Tanguy, Jacques Millo, Nicolas Passilly, Rodolphe Boudot

Journal of the Optical Society of America B · 2024 · 12 citations · 53 references

DOIFull text

Open access

Concepts

Abstract

We report on the development and short-term stability characterization of an optical frequency reference based on the spectroscopy of the rubidium two-photon transition at 778 nm in a microfabricated vapor cell. When compared against a 778 nm reference signal extracted from a frequency-doubled cavity-stabilized telecom laser, the short-term stability of the microcell frequency standard is 3.5×10 −13 τ −1/2 until 200 s, in good agreement with a phase noise level of +43dBrad 2 /Hz at 1 Hz offset frequency. The two main contributions to the short-term stability of the microcell reference are currently the photon shot noise and the intermodulation effect induced by the laser frequency noise. Retaining a relevant margin of progress, these results show the interest of this spectroscopic approach for the demonstration of high-stability miniaturized optical vapor cell clocks. Such clocks are poised to be highly beneficial for applications in navigation, communications, and metrology.

References