Validation of photoacoustic/ultrasound dual imaging in evaluating blood oxygen saturation

Sirui Liu, Rui Zhang, Tao Han, Yinhao Pan, Guangjie Zhang, Xing Long, Chenyang Zhao, Ming Wang, Xuelan Li, Fang Yang,

Biomedical Optics Express · 2022 · 22 citations · 30 references

DOIFull text

Open access

Concepts

Abstract

Photoacoustic imaging (PAI) was performed to evaluate oxygen saturation ( s O 2 ) of blood-mimicking phantoms, femoral arteries in beagles, and radial arteries in humans at various s O 2 plateaus. The accuracy (root mean square error, RMSE) of PAI s O 2 compared with reference s O 2 was calculated. In blood-mimicking phantoms, PAI achieved an accuracy of 1.49% and a mean absolute error (MAE) of 1.09% within 25 mm depth, and good linearity (R = 0.968; p < 0.001) was obtained between PAI s O 2 and reference s O 2 . In canine femoral arteries, PAI achieved an accuracy of 2.16% and an MAE of 1.58% within 8 mm depth (R = 0.965; p < 0.001). In human radial arteries, PAI achieved an accuracy of 3.97% and an MAE of 3.28% in depth from 4 to 14 mm (R = 0.892; p < 0.001). For PAI s O 2 evaluation at different depths in healthy volunteers, the RMSE accuracy of PAI s O 2 increased from 2.66% to 24.96% with depth increasing from 4 to 14 mm. Through the multiscale method, we confirmed the feasibility of the hand-held photoacoustic/ultrasound (PA/US) in evaluating s O 2 . These results demonstrate the potential clinical value of PAI in evaluating blood s O 2 . Consequently, protocols for verifying the feasibility of medical devices based on PAI may be established.

References

30