Metals · 2022 · 15 citations · 21 references
Natural SciencesChemistryAnalytical ChemistryElemental AnalysisMass Spectrometric AnalysisMass SpectrometryWaste CharacterizationGas FlowInductively Coupled Plasma Optical Emission SpectrometryInductively Coupled Plasma Optical Emission SpectroscopyMatrix Effect MitigationMicrowave DigestionMicrowave-assisted DigestionMicrowave-assisted Sample PreparationRecovery Test Analysis
In this article, inductively coupled plasma mass-spectrometry (ICP-MS) and inductively coupled plasma optical-emission spectrometry (ICP-OES) were used for the development of an analytical procedure for analysis of the waste of Sm-Co magnets. Experimental parameters related to microwave digestion processes and acid concentrations were optimized. Microwave digestion was carried out in mixtures of HF, HCl, HNO3 and H2SO4. The complete dissolution of the samples occurred in the system: 10 mL H2O, 2 mL HNO3, 10 mL HCl and 1 mL H2SO4. The dependence of the matrix effect on the ICP-MS analysis of waste Sm-Co magnets was studied and optimal instrumental parameters were investigated (nebulizer gas flow, sampling depth and potential at the extractor lens). The optimal conditions were a nebulizer gas flow of 0.85–0.90 L/min, a sampling depth of 101, potential at the extractor lens of −400 V and a sample flow rate of 50 rpm. A recovery test and inter-method experiments were performed to verify the accuracy of the proposed method.
21
Margareth Sugano Navarro, Sandra Andrade, H. Ulbrich et al. · Geostandards and Geoanalytical Research · 2008 · 173 citations · Full text
Nouri M. Hassan, Pat E. Rasmussen, Ewa Dąbek-Złotorzyńska et al. · Water Air & Soil Pollution · 2006 · 99 citations