Nuclear Engineering and Technology · 2020 · 16 citations · 10 references
EngineeringChemical EngineeringChemistryHeat TransferSensorsThermal AnalysisCalibrationGas SensorsPressure MeasurementPressure MetrologyPressure SensorsSensor CalibrationDensityPressure Sensor CharacterizationPressure Test CalibrationSurface Tension MeasurementThermal Expansion
A triple bubbler sensor was tested in LiCl–KCl molten salt from 450 to 525 °C in a transparent furnace to validate thermal-expansion corrections and provide additional molten salt data sets for calibration and validation of the sensor. In addition to these tests, a model was identified and further developed to accurately determine the density, surface tension, and depth from the measured bubble pressures. A unique feature of the model is that calibration constants can be estimated using independent depth measurements, which allow calibration and validation of the sensor in an electrorefiner where the salt density and surface tension are largely unknown. This model and approach were tested using the current and previous triple bubbler data sets, and results indicate that accuracies are as high as 0.03%, 4.6%, and 0.15% for density, surface tension, and depth, respectively.
10
George J. Janz · Journal of Physical and Chemical Reference Data · 1988 · 253 citations
Engineering-Scale Liquid Cadmium Cathode Experiments
D. Vaden, Shelly Li, B. R. Westphal et al. · Nuclear Technology · 2008 · 81 citations · Full text