Spirometry Measurement of Peak Inspiratory Flow Identifies Suboptimal Use of Dry Powder Inhalers in Ambulatory Patients with COPD

Alexander G. Duarte, Leon Tung, Wei Zhang, En Shuo Hsu, Yong‐Fang Kuo, Gulshan Sharma

Chronic Obstructive Pulmonary Diseases Journal of the COPD Foundation · 2019 · 54 citations · 17 references

DOIFull text

Open access

Concepts

Abstract

OBJECTIVES: device in ambulatory patients with COPD. METHODS: device and pulmonary function testing with calibrated equipment. Group characteristics and lung function were compared for patients with suboptimal (≤ 60 L/min) and optimal (> 60 L/min) PIFR. Receiver operating curve analysis determined the best maximal forced inspiratory flow (FIF max) value in identifying optimal PIFR by gender and height. RESULTS: <0.0001). Receiver operator curve analysis of FIF max to identify an optimal PIFR yielded an area under the curve of 0.79. Males < 65 inches had a suboptimal PIFR in 16.7 % of the male cohort, while females < 65 inches had a suboptimal PIFR in 27.4 % of the women. CONCLUSIONS: Suboptimal PIFR was present in 1 in 5 stable patients with COPD and was more frequent in short statured females. Spirometry determined FIF max was associated with PIFR based on gender and height.

References

17