EngineeringMechanical EngineeringInstrumentationSensorsVibrationMechanical MeasurementMechanical SensorsPiezoelectric MaterialsPressure SensorsSensor CalibrationVibration MeasurementAcceleration SensorsDynamic CalibrationDynamic CompensationEmpirical Mode DecompositionPiezoelectric Pressure SensingPiezoelectric Pressure SensorsPiezoelectric SensingPiezoelectric Sensors
Transient pressure testing is often accompanied by shock acceleration. Aiming at the acceleration-induced effects of pressure sensors, a dynamic compensation method combining empirical mode decomposition (EMD) with system identification theory (SIT) is proposed in this paper. This method is more effective at reducing the error of the acceleration-induced effects without affecting the sensor's sensitivity and inherent frequency. The principle and theoretical basis of acceleration-induced effects is analyzed, and the static and dynamic acceleration-induced effects on the quartz crystal of a piezoelectric pressure sensor are performed. An acceleration-induced effects dynamic calibration system is built using a Machete hammer, which generates acceleration signals with larger amplitude and narrower pulse width, and an autoregressive exogenous (ARX)mathematical model of acceleration-induced effects is obtained using empirical mode decomposition-system identification theory (EMD-SIT). A digital compensation filter for acceleration-induced effects is designed on the basis of this model. Experimental results explain that the acceleration-induced effects of the pressure sensor were less than 11% after using the digital compensation filter. A series of test data verify the accuracy, reliability, and generality of the model.
10