Publication | Open Access
Field evaluation of quantitative point of care diagnostics to measure glucose-6-phosphate dehydrogenase activity
86
Citations
9
References
2018
Year
BS and SG are the first quantitative diagnostics to measure G6PD activity reliably at the bedside and represent suitable alternatives to spectrophotometry in resource poor settings. If samples are stored at 4°C, G6PD activity can be measured reliably for at least 7 days after sample collection.